Screw- vs cement-implant-retained restorations: an experimental study in the Beagle. Part 1. Screw and abutment loosening

The causes of implant failures can be biological or mechanical. The mechanical causes include fracture of the implant, fracture of the abutment, and loosening of the abutment. Numerous studies show that abutment loosening constitutes one of the marked implant postsurgery complications requiring clin...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of oral implantology 2005, Vol.31 (5), p.242-246
Hauptverfasser: Assenza, Bartolomeo, Scarano, Antonio, Leghissa, Giulio, Carusi, Giorgio, Thams, Ulf, Roman, Fidel San, Piattelli, Adriano
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The causes of implant failures can be biological or mechanical. The mechanical causes include fracture of the implant, fracture of the abutment, and loosening of the abutment. Numerous studies show that abutment loosening constitutes one of the marked implant postsurgery complications requiring clinical intervention. The aim of the present study was to evaluate the incidence of the screw loosening in screwed or cemented abutments. Six adult male Beagles were used. In each dog, the first molars and 2 premolars were extracted. The sutures were removed after 7 days. After 3 months, 10 implants were placed in each dog, 5 in the right mandible and 5 in the left mandible. The abutments either were screwed in (n=30) by applying a total strength of 30 N/cm or were cemented (n=30). After 12 months, 8 (27%) loosened screws were present in screwed abutments, whereas no abutment loosening was observed in cemented abutments (P = .0001). Screwed abutments are often submitted to nonaxial loads that determine screw and abutment loosening.
ISSN:0160-6972