Analysis of 18F-labelled synthesis products on TLC plates: Comparison of radioactivity scanning, film autoradiography, and a phosphoimaging technique
We compared radioactivity scanning, film autoradiography, and digital photostimulated luminescence (PSL) autoradiography (phosphoimaging technique) in detection of radioactivity on thin-layer chromatography (TLC) plates. TLC combined with radioactivity detection is rapid, simple, and relatively flex...
Gespeichert in:
Veröffentlicht in: | Applied radiation and isotopes 2006-09, Vol.64 (9), p.1043-1047 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We compared radioactivity scanning, film autoradiography, and digital photostimulated luminescence (PSL) autoradiography (phosphoimaging technique) in detection of radioactivity on thin-layer chromatography (TLC) plates. TLC combined with radioactivity detection is rapid, simple, and relatively flexible. Here,
18F-labelled synthesis products were analyzed by TLC and the radioactivity distribution on the plates determined using the three techniques. Radioactivity scanning is appropriate only with good chromatographic resolution and previously validated scanning parameters. Film autoradiography exhibits poor linearity if radioactivity varies greatly. PSL provides high sensitivity and resolution and superior linearity compared with the other methods. |
---|---|