Right Heart Failure Due to Loss of Right Ventricular Capture in a Patient with Atrioventricular Junction Ablation and Biventricular Pacing

We describe the case of a patient with atrioventricular (AV) junction ablation and chronic biventricular pacing in which intermittent dysfunction of the right ventricular (RV) lead resulted in left ventricular (LV) stimulation alone and onset of severe right heart failure. Restoration of biventricul...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 2005-10, Vol.28 (10), p.1127-1130
Hauptverfasser: RAFFA, SANTI, FANTONI, CECILIA, RESTAURI, LUIGIA, AURICCHIO, ANGELO
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We describe the case of a patient with atrioventricular (AV) junction ablation and chronic biventricular pacing in which intermittent dysfunction of the right ventricular (RV) lead resulted in left ventricular (LV) stimulation alone and onset of severe right heart failure. Restoration of biventricular pacing by increasing device output and then performing lead revision resolved the issue. This case provides evidence that LV pacing alone in patients with AV junction ablation may lead to severe right heart failure, most likely as a result of iatrogenic mechanical dyssynchrony within the RV. Thus, probably this pacing mode should be avoided in pacemaker‐dependent patients with heart failure.
ISSN:0147-8389