The Effect of Total Hip Arthroplasty Cup Design on Polyethylene Wear Rate
Using 743 total hip arthroplasties that included 6 hemispheric porous-coated cup designs, this study used a multiple linear regression to identify those factors that influenced polyethylene wear rates. Wear rates for each hip were based on serial head penetration measurement made with computer-assis...
Gespeichert in:
Veröffentlicht in: | The Journal of arthroplasty 2005-10, Vol.20 (7 Suppl 3), p.63-72 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Using 743 total hip arthroplasties that included 6 hemispheric porous-coated cup designs, this study used a multiple linear regression to identify those factors that influenced polyethylene wear rates. Wear rates for each hip were based on serial head penetration measurement made with computer-assisted techniques. Implant factors associated with an increased wear rate included terminal sterilization with a non–cross-linking chemical surface treatment, a 4-mm lateralized liner, a cobalt-chrome femoral head, and a longer shelf life for liners
γ-irradiated in air. After accounting for these implant characteristics and patient factors, wear rates among the 6 cup designs were not significantly different (
P = .89). Although polyethylene wear is frequently characterized for specific implant designs, this study demonstrated that there are several common factors that influence polyethylene wear rates. |
---|---|
ISSN: | 0883-5403 1532-8406 |
DOI: | 10.1016/j.arth.2005.05.007 |