Hylamer wear rates and shelf life: a clinical correlation

We evaluated the performance of 84 Hylamer polyethylene components and zirconia modular femoral heads which were implanted in a single institution over a 5 year period. Patients were followed up for a mean of 6.2 years. The mean rate of penetration was 0.58 mm/year (0.05-1.0). Lot/batch numbers were...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Acta orthopaedica belgica 2005-08, Vol.71 (4), p.429-434
Hauptverfasser: Kiely, Paul D, Harty, James A, McElwain, John P
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We evaluated the performance of 84 Hylamer polyethylene components and zirconia modular femoral heads which were implanted in a single institution over a 5 year period. Patients were followed up for a mean of 6.2 years. The mean rate of penetration was 0.58 mm/year (0.05-1.0). Lot/batch numbers were available in the clinical case notes of 54 patients. Using the lot/batch numbers, we were able to determine the Hylamer cups' shelf lifes. We found that liners with a shelf life greater than 10 months had a significantly greater linear wear (0.38 mm/year) than those implanted in less than 10 months (0.05 mm/year).
ISSN:0001-6462