Place-pitch discrimination of single- versus dual-electrode stimuli by cochlear implant users
Simultaneous or near-simultaneous activation of adjacent cochlear implant electrodes can produce pitch percepts intermediate to those produced by each electrode separately, thereby increasing the number of place-pitch steps available to cochlear implant listeners. To estimate how many distinct pitch...
Gespeichert in:
Veröffentlicht in: | The Journal of the Acoustical Society of America 2005-08, Vol.118 (2), p.623-626 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Simultaneous or near-simultaneous activation of adjacent cochlear implant electrodes can produce pitch percepts intermediate to those produced by each electrode separately, thereby increasing the number of place-pitch steps available to cochlear implant listeners. To estimate how many distinct pitches could be generated with simultaneous dual-electrode stimulation, the present study measured place-pitch discrimination thresholds for single- versus dual-electrode stimuli in users of the Clarion CII device. Discrimination thresholds were expressed as the proportion of current directed to the secondary electrode of the dual-electrode pair. For 16 of 17 electrode pairs tested in six subjects, thresholds ranged from 0.11 to 0.64, suggesting that dual-electrode stimuli can produce 2-9 discriminable pitches between the pitches of single electrodes. Some subjects demonstrated a level effect, with better place-pitch discrimination at higher stimulus levels. Equal loudness was achieved with dual-electrode stimuli at net current levels that were similar to or slightly higher than those for single-electrode stimuli. |
---|---|
ISSN: | 0001-4966 1520-8524 |
DOI: | 10.1121/1.1937362 |