Comparative evaluation of TRI-DOT Rapid HIV test with fourth-generation ELISA for the detection of human immunodeficiency virus

This study evaluated the TRI-DOT Rapid HIV test for the early detection of human immunodeficiency virus (HIV) infection in comparison with a fourth-generation ELISA (Vironostika HIV Uniform II). Of 23 609 sera, seven (0.03%) gave discordant results. Six of these were reactive only by the fourth-gene...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical microbiology and infection 2005-10, Vol.11 (10), p.850-852
Hauptverfasser: Sudha, T., Teja, V.D., Gopal, M., Rajesh, M., Lakshmi, V.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This study evaluated the TRI-DOT Rapid HIV test for the early detection of human immunodeficiency virus (HIV) infection in comparison with a fourth-generation ELISA (Vironostika HIV Uniform II). Of 23 609 sera, seven (0.03%) gave discordant results. Six of these were reactive only by the fourth-generation assay and were p24 antigen-positive by VIDAS DUO, Western blot and qualitative RT-PCR tests. The remaining discordant serum was considered to be false-positive by the TRI-DOT assay, as it was negative by repeat ELISA and Western blot tests. The sensitivity and specificity of the TRI-DOT test were 99.48% and 99.99%, respectively, compared with the fourth-generation ELISA.
ISSN:1198-743X
1469-0691
DOI:10.1111/j.1469-0691.2005.01231.x