Effects of filtering and delays on the estimates of a nonlinear respiratory mechanics model
Estimation of mechanical properties of the respiratory system may be disturbed by instrumentation and physical set-up. The effects of lowpass filtering, filter mismatch and inter-channel delay in the digital converter are assessed on numerically simulated signals from a nonlinear model of the respir...
Gespeichert in:
Veröffentlicht in: | Respiratory physiology & neurobiology 2005-10, Vol.148 (3), p.309-314 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Estimation of mechanical properties of the respiratory system may be disturbed by instrumentation and physical set-up. The effects of lowpass filtering, filter mismatch and inter-channel delay in the digital converter are assessed on numerically simulated signals from a nonlinear model of the respiratory system. Large biases in model parameter estimates (up to about −300% for some parameters) were caused by these instrumental interferences and were reduced by including an inertance in the retrieved model. The results reinforce the importance of a careful evaluation of the instrumental set-up used in physiological measurements. |
---|---|
ISSN: | 1569-9048 1878-1519 |
DOI: | 10.1016/j.resp.2005.02.009 |