Effects of filtering and delays on the estimates of a nonlinear respiratory mechanics model

Estimation of mechanical properties of the respiratory system may be disturbed by instrumentation and physical set-up. The effects of lowpass filtering, filter mismatch and inter-channel delay in the digital converter are assessed on numerically simulated signals from a nonlinear model of the respir...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Respiratory physiology & neurobiology 2005-10, Vol.148 (3), p.309-314
Hauptverfasser: Jandre, Frederico C., Carvalho, Alysson Roncally Silva, Pino, Alexandre Visintainer, Giannella-Neto, Antonio
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Estimation of mechanical properties of the respiratory system may be disturbed by instrumentation and physical set-up. The effects of lowpass filtering, filter mismatch and inter-channel delay in the digital converter are assessed on numerically simulated signals from a nonlinear model of the respiratory system. Large biases in model parameter estimates (up to about −300% for some parameters) were caused by these instrumental interferences and were reduced by including an inertance in the retrieved model. The results reinforce the importance of a careful evaluation of the instrumental set-up used in physiological measurements.
ISSN:1569-9048
1878-1519
DOI:10.1016/j.resp.2005.02.009