High-flow nasal cannula as a device to provide continuous positive airway pressure in infants

To measure the intrapharyngeal pressure (IPP) generated by high-flow nasal cannula (HFNC) at varying flow rates up to 5 l min(-1). We studied 14 infants on HFNC or NCPAP using IPP manometry to measure the IPP generated at flows of 1, 2, 3, 4 and 5 l min(-1). On average flow rates of 1, 2, 3, 4 and 5...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of Perinatology 2007-12, Vol.27 (12), p.772-775
Hauptverfasser: Spence, K L, Murphy, D, Kilian, C, McGonigle, R, Kilani, R A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To measure the intrapharyngeal pressure (IPP) generated by high-flow nasal cannula (HFNC) at varying flow rates up to 5 l min(-1). We studied 14 infants on HFNC or NCPAP using IPP manometry to measure the IPP generated at flows of 1, 2, 3, 4 and 5 l min(-1). On average flow rates of 1, 2, 3, 4 and 5 l min(-1) generated an IPP (cm of H(2)O) of 1.70+/-0.34, 1.75+/-0.2, 2.62+/-0.28, 3.78+/-0.44 and 4.84+/-0.51 respectively. HFNC delivers significant IPP and is potentially a well-tolerated and viable option to provide CPAP at flows of > or = 3 min(-1) in infants with respiratory distress.
ISSN:0743-8346
1476-5543
DOI:10.1038/sj.jp.7211828