High-flow nasal cannula as a device to provide continuous positive airway pressure in infants
To measure the intrapharyngeal pressure (IPP) generated by high-flow nasal cannula (HFNC) at varying flow rates up to 5 l min(-1). We studied 14 infants on HFNC or NCPAP using IPP manometry to measure the IPP generated at flows of 1, 2, 3, 4 and 5 l min(-1). On average flow rates of 1, 2, 3, 4 and 5...
Gespeichert in:
Veröffentlicht in: | Journal of Perinatology 2007-12, Vol.27 (12), p.772-775 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To measure the intrapharyngeal pressure (IPP) generated by high-flow nasal cannula (HFNC) at varying flow rates up to 5 l min(-1).
We studied 14 infants on HFNC or NCPAP using IPP manometry to measure the IPP generated at flows of 1, 2, 3, 4 and 5 l min(-1).
On average flow rates of 1, 2, 3, 4 and 5 l min(-1) generated an IPP (cm of H(2)O) of 1.70+/-0.34, 1.75+/-0.2, 2.62+/-0.28, 3.78+/-0.44 and 4.84+/-0.51 respectively.
HFNC delivers significant IPP and is potentially a well-tolerated and viable option to provide CPAP at flows of > or = 3 min(-1) in infants with respiratory distress. |
---|---|
ISSN: | 0743-8346 1476-5543 |
DOI: | 10.1038/sj.jp.7211828 |