HPLC–MS method for the simultaneous quantification of the new HIV protease inhibitor darunavir, and 11 other antiretroviral agents in plasma of HIV-infected patients
A new method using high performance liquid chromatography coupled with electrospray mass spectrometry (HPLC–MS) was developed and validated, for the quantification of plasma concentration of the new protease inhibitors darunavir (DRV) and other 11 antiretroviral agents (ritonavir, amprenavir, atazan...
Gespeichert in:
Veröffentlicht in: | Journal of chromatography. B, Analytical technologies in the biomedical and life sciences Analytical technologies in the biomedical and life sciences, 2007-11, Vol.859 (2), p.234-240 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new method using high performance liquid chromatography coupled with electrospray mass spectrometry (HPLC–MS) was developed and validated, for the quantification of plasma concentration of the new protease inhibitors darunavir (DRV) and other 11 antiretroviral agents (ritonavir, amprenavir, atazanavir, lopinavir, saquinavir, indinavir, nelfinavir and its metabolite M-8, nevirapine, efavirenz and tipranavir). A simple protein precipitation extraction procedure was applied on 50
μl of plasma aliquots and chromatographic separation of drugs and Internal Standard (quinoxaline) was achieved with a gradient (acetonitrile and water with formic acid 0.05%) on an C-18 reverse phase analytical column with 25
min of analytical run. Calibration curves were optimised according to expected ranges of drug concentrations in patients, and correlation coefficient (
r
2) was higher than 0.998 for all analytes. Mean intra- and inter-day precision (relative standard deviation %) for all compounds were 8.4 and 8.3%, respectively, and mean accuracy (% of deviation from nominal level) was 3.9%. Extraction recovery ranged within 93 and 105% for all drugs analysed. This novel HPLC–MS methodology allows a specific, sensitive and reliable determination of DRV and 11 other antiretrovirals. In our hand, it was used to measure DRV and ritonavir plasma concentration in HIV-positive patients, and it is now successfully applied for routine therapeutic drug monitoring and pharmacokinetics studies. |
---|---|
ISSN: | 1570-0232 1873-376X |
DOI: | 10.1016/j.jchromb.2007.10.003 |