High-performance liquid chromatography method for quantifying sphingomyelin in rat brain

A rapid, reproducible and accurate high-performance liquid chromatographic (HPLC) method for the quantitative determination of sphingomyelin in rat brain was developed and validated using normal-phase silica gel column, acetonitrile–methanol–water (65:18:17 (v/v)) at a flow rate of 1 ml/min, isocrat...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of chromatography. B, Analytical technologies in the biomedical and life sciences Analytical technologies in the biomedical and life sciences, 2007-11, Vol.859 (1), p.131-136
Hauptverfasser: Azzam, Diana J., Usta, Julnar A., Mouneimne, Youssef, Hokayem, Jimmy A. El, Mikati, Mohamad A.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A rapid, reproducible and accurate high-performance liquid chromatographic (HPLC) method for the quantitative determination of sphingomyelin in rat brain was developed and validated using normal-phase silica gel column, acetonitrile–methanol–water (65:18:17 (v/v)) at a flow rate of 1 ml/min, isocratic elution, UV detection at 207 nm and 1,2-dimyristoyl- sn-glycero-3-phosphocholine as an internal standard. Total run time was 10.0 min. The calibration curve was linear over the range of 0.025–0.4 mg/ml sphingomyelin ( R 2 > 0.99). The intra-day coefficient of variation ranged from 1.4% to 2.2%. The average inter-day coefficient of variation over a period of 4 days was 3.1%. The practical limit of detection was 0.005 mg/ml with a quantification limit of 0.01 mg/ml.
ISSN:1570-0232
1873-376X
DOI:10.1016/j.jchromb.2007.09.005