Prostate-specific antigen immunosensing based on mixed self-assembled monolayers, camel antibodies and colloidal gold enhanced sandwich assays
Prostate-specific antigen (PSA) is a valuable biomarker for prostate cancer screening. We developed a PSA immunoassay on a commercially available surface plasmon resonance biosensor. Our PSA receptor molecule consists of a single domain antigen-binding fragment, cAbPSA-N7, derived from dromedary hea...
Gespeichert in:
Veröffentlicht in: | Biosensors & bioelectronics 2005-09, Vol.21 (3), p.483-490 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Prostate-specific antigen (PSA) is a valuable biomarker for prostate cancer screening. We developed a PSA immunoassay on a commercially available surface plasmon resonance biosensor. Our PSA receptor molecule consists of a single domain antigen-binding fragment, cAbPSA-N7, derived from dromedary heavy-chain antibodies and identified after phage display. It binds PSA with a high
k
on value of 1.9
×
10
6
M
−1
s
−1, and was covalently immobilised on a gold substrate via a mixed self-assembled monolayer (SAM) of alkanethiols by using carbodiimide-coupling chemistry in 10
mM acetate buffer pH 5.5 to obtain an optimal pre-concentration. The best performing and optimised mixed SAM consisted of (10%) 16-mercapto-1-hexadecanoic acid (16-MHA) for covalent cAbPSA-N7 immobilisation and (90%) 11-mercapto-1-undecanol (11-MUOH) to minimise non-specific adsorption of the analyte. In this way, two advantages are incorporated in a single coupling layer. Up to 28
fmol/mm
2 of cAbPSA-N7 could be immobilised and 30% of its binding sites participate actively in PSA interaction. In addition, the optimised layer showed also optimal performance to assess physiological samples. Although PSA concentrations as low as 10
ng/ml could be detected directly, this detection limit could be enhanced to PSA levels in the sub ng/ml range by introducing a sandwich assay involving a biotinylated secondary antibody and streptavidin modified gold nanoparticles. This approach realises the PSA detection at clinical relevant concentrations. |
---|---|
ISSN: | 0956-5663 1873-4235 |
DOI: | 10.1016/j.bios.2004.11.016 |