Control of tracheal cuff pressure : a pilot study using a pneumatic device

To evaluate the efficacy of a simple mechanical device to maintain constant endotracheal cuff pressure (Pcuff) during mechanical ventilation (large encased inflatable cuff connected to the endotracheal cuff and receiving constant pressure from a heavy mass attached to an articulated arm). Single-cen...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Intensive care medicine 2007-01, Vol.33 (1), p.128-132
Hauptverfasser: DUGUET, Alexandre, D'AMICO, Leda, BIONDI, Giuseppina, PRODANOVIC, Helene, GONZALEZ-BERMEJO, Jésus, SIMILOWSKI, Thomas
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To evaluate the efficacy of a simple mechanical device to maintain constant endotracheal cuff pressure (Pcuff) during mechanical ventilation (large encased inflatable cuff connected to the endotracheal cuff and receiving constant pressure from a heavy mass attached to an articulated arm). Single-center, prospective, randomized, crossover, pilot study in a medical intensive care unit. Nine consecutive mechanically ventilated patients (age 62+/-20 years, SAPS II score 39+/-15). Control day: Pcuff monitored and adjusted with a manometer (Hi-Lo, Tyco Healthcare) according to current recommendations (twice a day and after each intervention on the tracheal tube); initial target Pcuff 22-28 cmH20. Prototype day: test device connected to the endotracheal cuff; same initial target. Continuous Pcuff recording during both days. Control and prototype days in random order. Pcuff values over 50 cmH20 were recorded in six patients during the control day (178+/-159min), never during the prototype day. During the control day, Pcuff was between 30 and 50 cmH20 for 29+/-25% of the time, vs 0.3+/-0.3% during the prototype day (p
ISSN:0342-4642
1432-1238
DOI:10.1007/s00134-006-0417-x