Selective detection of extracellular glutamate in brain tissue using microelectrode arrays coated with over-oxidized polypyrrole
We describe the selective measurement of extracellular glutamate concentration in rodent brain using ceramic-based platinum microelectrode arrays (MEAs) coated with electropolymerized, over-oxidized polypyrrole (OPP) as a permselective barrier to interference from dopamine and ascorbate. OPP-coated...
Gespeichert in:
Veröffentlicht in: | Analyst (London) 2007-11, Vol.132 (11), p.1107-1111 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We describe the selective measurement of extracellular glutamate concentration in rodent brain using ceramic-based platinum microelectrode arrays (MEAs) coated with electropolymerized, over-oxidized polypyrrole (OPP) as a permselective barrier to interference from dopamine and ascorbate. OPP-coated MEAs displayed similar sensitivity (97 +/- 9 nA microM(-1) cm(-2)) and response time (ca. 1 s) to glutamate as previously described Nafion-coated MEAs but, unlike Nafion-coated MEAs, were unresponsive to 5 microM dopamine. Such OPP-coated MEAs afforded selective detection of evoked glutamate release in the dopamine-rich striatum of the mouse brain. |
---|---|
ISSN: | 0003-2654 1364-5528 |
DOI: | 10.1039/b706880h |