A 10-year comparison of explanted Hancock-II and Carpentier–Edwards supraannular bioprostheses

Abstract Background Bioprosthetic heart valves are more frequently being used in valve replacement procedures today. Although second-generation bioprosthetic valves have improved functionality over their first-generation counterparts, they still often fail due to primary tissue degeneration. Methods...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Cardiovascular pathology 2007, Vol.16 (1), p.4-13
Hauptverfasser: Butany, Jagdish, Leong, Shaun W, Cunningham, Kristopher S, D'Cruz, Geoffrey, Carmichael, Khenan, Yau, Terrence M
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract Background Bioprosthetic heart valves are more frequently being used in valve replacement procedures today. Although second-generation bioprosthetic valves have improved functionality over their first-generation counterparts, they still often fail due to primary tissue degeneration. Methods This study examines two second-generation porcine valves after surgical explantation, the Hancock-II (HAN; Medtronic Heart Valve Division, Irvine, CA, USA) and the Carpentier–Edwards supraannular (CE-SAV; Baxter Healthcare Corporation, now Edwards LifeSciences, Irvine, CA, USA), with special attention to morphological/histological changes and reasons for valve failure. A total of 98 HAN and 65 CE-SAV valves were explanted and seen over a 10-year period. Results CE-SAV valves had a longer average implant duration than HAN valves (13.9±3.9 years vs. 10.0±5.1 years). Compared with HAN valves, CE-SAV valves also had a higher incidence of stent deformation (41.5% vs. 14.3%), calcification (75.4% vs. 54.1%), and pannus (100% vs. 91.8%). Conclusions The greater degenerative changes seen with CE-SAV valves over HAN valves may be due to the longer implant duration of CE-SAV valves in this series. To our knowledge, the present study is the first direct morphological comparison of these two valve models.
ISSN:1054-8807
1879-1336
DOI:10.1016/j.carpath.2006.06.003