Late Aneurysm Rupture after Repressurization of a Thrombosed Stent-Graft
Purpose: To report late abdominal aortic aneurysm (AAA) rupture after endovascular stent-graft repair despite complete thrombotic stent-graft occlusion. Case Report: A 65-year-old man underwent successful endovascular aneurysm repair (EVAR) with a Stentor device in 1995. In the interim course, the p...
Gespeichert in:
Veröffentlicht in: | Journal of endovascular therapy 2007-10, Vol.14 (5), p.672-675 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose:
To report late abdominal aortic aneurysm (AAA) rupture after endovascular stent-graft repair despite complete thrombotic stent-graft occlusion.
Case Report:
A 65-year-old man underwent successful endovascular aneurysm repair (EVAR) with a Stentor device in 1995. In the interim course, the patient developed complete thrombotic stent-graft occlusion, which was treated with an axillobifemoral bypass. After 8 years, the patient presented with a reperfused and ruptured infrarenal AAA. Open repair was performed, with a good clinical result and exclusion of the AAA.
Conclusion:
Thrombosed stent-grafts and aneurysms can transmit systemic arterial pressure and cause late rupture. Lifelong surveillance is mandatory in EVAR patients. |
---|---|