Late Aneurysm Rupture after Repressurization of a Thrombosed Stent-Graft

Purpose: To report late abdominal aortic aneurysm (AAA) rupture after endovascular stent-graft repair despite complete thrombotic stent-graft occlusion. Case Report: A 65-year-old man underwent successful endovascular aneurysm repair (EVAR) with a Stentor device in 1995. In the interim course, the p...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of endovascular therapy 2007-10, Vol.14 (5), p.672-675
Hauptverfasser: Geisbüsch, Philipp, Schumacher, Hardy, Hansmann, Jochen, Allenberg, Jens-Rainer, Böckler, Dittmar
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose: To report late abdominal aortic aneurysm (AAA) rupture after endovascular stent-graft repair despite complete thrombotic stent-graft occlusion. Case Report: A 65-year-old man underwent successful endovascular aneurysm repair (EVAR) with a Stentor device in 1995. In the interim course, the patient developed complete thrombotic stent-graft occlusion, which was treated with an axillobifemoral bypass. After 8 years, the patient presented with a reperfused and ruptured infrarenal AAA. Open repair was performed, with a good clinical result and exclusion of the AAA. Conclusion: Thrombosed stent-grafts and aneurysms can transmit systemic arterial pressure and cause late rupture. Lifelong surveillance is mandatory in EVAR patients.
ISSN:1526-6028
1545-1550