Synchrotron radiation-based irradiance calibration from 200 to 400 nm at the Synchrotron Ultraviolet Radiation Facility III
A new facility for measuring irradiance in the UV was commissioned recently at the National Institute of Standards and Technology (NIST). The facility uses the calculable radiation from the Synchrotron Ultraviolet Radiation Facility as the primary standard. To measure the irradiance from a source un...
Gespeichert in:
Veröffentlicht in: | Applied Optics 2007-01, Vol.46 (1), p.25-35 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new facility for measuring irradiance in the UV was commissioned recently at the National Institute of Standards and Technology (NIST). The facility uses the calculable radiation from the Synchrotron Ultraviolet Radiation Facility as the primary standard. To measure the irradiance from a source under test, an integrating sphere spectrometer-detector system measures both the source under test and the synchrotron radiation sequentially, and the irradiance from the source under test can be determined. In particular, we discuss the calibration of deuterium lamps using this facility from 200 to 400 nm. This facility improves the current NIST UV irradiance scale to a relative measurement uncertainty of 1.2% (k=2). |
---|---|
ISSN: | 1559-128X 0003-6935 1539-4522 |
DOI: | 10.1364/AO.46.000025 |