Analysis of perchlorate in foods and beverages by ion chromatography coupled with tandem mass spectrometry (IC-ESI-MS/MS)
A new IC-ESI-MS/MS method, with simple sample preparation procedure, has been developed for quantification and confirmation of perchlorate (ClO 4 −) anions in water, fresh and canned food, wine and beer samples at low part-per-trillion (ng l −1) levels. To the best of our knowledge, this is the firs...
Gespeichert in:
Veröffentlicht in: | Analytica chimica acta 2006-05, Vol.567 (1), p.39-47 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A new IC-ESI-MS/MS method, with simple sample preparation procedure, has been developed for quantification and confirmation of perchlorate (ClO
4
−) anions in water, fresh and canned food, wine and beer samples at low part-per-trillion (ng
l
−1) levels. To the best of our knowledge, this is the first time an analytical method is used for determination of perchlorate in wine and beer samples. The IC-ESI-MS/MS instrumentation consisted of an ICS-2500 ion chromatography (IC) system coupled to either an API 2000™ or an API 3200™ mass spectrometer. The IC-ESI-MS/MS system was optimized to monitor two pairs of precursor and fragment ion transitions, i.e., multiple reaction monitoring (MRM). All samples had oxygen-18 isotope labeled perchlorate internal standard (ISTD) added prior to extraction. Chlorine isotope ratio (
35Cl/
37Cl) was used as a confirmation tool. The transition of
35Cl
16O
4
− (
m/
z 98.9) into
35Cl
16O
3
− (
m/
z 82.9) was monitored for quantifying the main analyte; the transition of
37Cl
16O
4
− (
m/
z 100.9) into
37Cl
16O
3
− (
m/
z 84.9) was monitored for examining a proper isotopic abundance ratio of
35Cl/
37Cl; and the transition of
35Cl
18O
4
− (
m/
z 107.0) into
35Cl
18O
3
− (
m/
z 89.0) was monitored for quantifying the internal standard. The minimum detection limit (MDL) for this method in de-ionized water is 5
ng
l
−1 (ppt) using the API 2000™ mass spectrometer and 0.5
ng
l
−1 using the API 3200™ mass spectrometer. Over 350 food and beverage samples were analyzed mostly in triplicate. Except for four, all samples were found to contain measurable amounts of perchlorate. The levels found ranged from 5
ng
l
−1 to 463.5
±
6.36
μg
kg
−1 using MRM 98.9
→
82.9 and 100
μl injection. |
---|---|
ISSN: | 0003-2670 1873-4324 |
DOI: | 10.1016/j.aca.2006.03.012 |