Quantification of Art v 1 and Act c 1 being major allergens of mugwort pollen and kiwi fruit extracts in mass-units by ion-exchange HPLC-UV method
A simple ion-exchange HPLC-UV method was developed for determination of major allergens from mugwort pollen and kiwi fruit extracts in mass-units. The separation of Art v 1 and Act c 1 from other components in the extracts was achieved in one step. The extinction coefficients used in the study were...
Gespeichert in:
Veröffentlicht in: | Journal of chromatography. B, Analytical technologies in the biomedical and life sciences Analytical technologies in the biomedical and life sciences, 2007-10, Vol.857 (2), p.188-194 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A simple ion-exchange HPLC-UV method was developed for determination of major allergens from mugwort pollen and kiwi fruit extracts in mass-units. The separation of Art v 1 and Act c 1 from other components in the extracts was achieved in one step. The extinction coefficients used in the study were theoretically determined and compared to the extinction coefficients determined by gravimetry. We also reported a close correlation of the major allergen contents with the overall allergenic potency of the extracts determined by inhibition ELISA. This method could be a useful tool for standardization of allergenic extracts for clinical use. |
---|---|
ISSN: | 1570-0232 1873-376X |
DOI: | 10.1016/j.jchromb.2007.07.015 |