Radiostereometric Analysis Comparison of Wear of Highly Cross-Linked Polyethylene Against 36- vs 28-mm Femoral Heads
Abstract This study used radiostereometric analysis (RSA) to compare the femoral head penetration of 28- vs 36-mm-diameter femoral heads into highly cross-linked polyethylene in 2 groups of total hip arthroplasty patients. Thirty patients were enrolled in this RSA study using highly cross-linked pol...
Gespeichert in:
Veröffentlicht in: | The Journal of arthroplasty 2007-09, Vol.22 (6), p.125-129 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract This study used radiostereometric analysis (RSA) to compare the femoral head penetration of 28- vs 36-mm-diameter femoral heads into highly cross-linked polyethylene in 2 groups of total hip arthroplasty patients. Thirty patients were enrolled in this RSA study using highly cross-linked polyethylene (Longevity, Zimmer Inc, Warsaw, Idaho) against either 28- or 36-mm-diameter cobalt chrome femoral heads. At 3-year follow-up, there was no significant difference in the total average femoral head penetration, including both creep and wear, using 3 methods of RSA measurement between the 2 groups. Importantly, after bedding-in, there was no further significant increase in the amount of femoral head penetration (ie, wear) with either head size between years 1 and 3. There were no radiographic signs of lysis or radiolucencies at a minimum 3-year follow-up. |
---|---|