Quantification of Left Ventricular Volumes and Function in Patients with Cardiomyopathies by Real-time Three-dimensional Echocardiography: A Head-to-Head Comparison Between Two Different Semiautomated Endocardial Border Detection Algorithms
Objective We evaluated two different commercially available real-time 3-dimensional echocardiographic semiautomated border detection algorithms for left ventricular (LV) volume analysis in patients with cardiomyopathy and distorted LV geometry. Methods A total of 53 patients in sinus rhythm with var...
Gespeichert in:
Veröffentlicht in: | Journal of the American Society of Echocardiography 2007-09, Vol.20 (9), p.1042-1049 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objective We evaluated two different commercially available real-time 3-dimensional echocardiographic semiautomated border detection algorithms for left ventricular (LV) volume analysis in patients with cardiomyopathy and distorted LV geometry. Methods A total of 53 patients in sinus rhythm with various types of cardiomyopathy (mean age 56 ± 11 years, 28 men) and adequate 2-dimensional image quality were included. The real-time 3-dimensional echocardiographic multiplane interpolation (MI) and full volume reconstruction (FVR) methods were used for LV volume analysis. Magnetic resonance imaging was used as the reference method. Results A strong correlation ( R2 > 0.95) was found for all LV volume and ejection fraction measurements by either real-time 3-dimensional echocardiographic method. Analysis time was shorter with the FVR method (6 ± 2 vs 15 ± 4 minutes, P < .01) as compared with the MI method. Bland-Altman analysis showed greater underestimation of end-diastolic and end-systolic volumes by MI compared with FVR. For the MI method a bias of −24.0 mL (−15.0% of the mean) for end-diastolic volume and −11.3 mL (−18.0% of the mean) for end-systolic volume was found. For FVR analysis these values were −9.9 mL (−6.0% of the mean) and −5.0 mL (−9.0% of the mean), respectively. Ejection fraction was similar for the MI and FVR method with a mean difference compared with magnetic resonance imaging of 0.6 (1.0%) and 0.8 (1.3%), respectively. Conclusion In patients with cardiomyopathy, distorted LV geometry, and good 2-dimensional image quality, the FVR method is faster and more accurate than the MI method in assessment of LV volumes. |
---|---|
ISSN: | 0894-7317 1097-6795 |
DOI: | 10.1016/j.echo.2007.02.011 |