Validation of antibody screening by indirect antigloblin test and ABO blood typing by filtration and microplate techniques: assessment of robustness

According to requirements of the French Committee for Accreditation (Cofrac), it is essential to use validated and standardised methods in Immunohematology. This imposes first the knowledge of metrological tolerances for all the technics. Two multicenter studies were carried out to define the maxima...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Transfusion clinique et biologique : journal de la Société française de transfusion sanguine 2006-10, Vol.13 (4), p.271-277
Hauptverfasser: Mannessier, L, Delamaire, M, Bouix, O, Krause, C, Roubinet, F
Format: Artikel
Sprache:fre
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:According to requirements of the French Committee for Accreditation (Cofrac), it is essential to use validated and standardised methods in Immunohematology. This imposes first the knowledge of metrological tolerances for all the technics. Two multicenter studies were carried out to define the maximal acceptable deviations concerning incubation temperature and time, volumes of patient plasma and tests cells for antibody screening using indirect antiglobulin test on one hand and for reverse grouping on another hand. All equipment used (temperature test chamber, chronometer, pipettes) were calibrated according to Cofrac standards. The antibody screenings were performed manually using 3 different filtration systems: ID Diamed, Biovue Ortho and Scangel Biorad, the same tests cells, a standard 20 ng/mL anti RH1, a positive control anti KEL1 and a negative control; the reverse blood grouping was performed manually using the above mentionned filtration systems and microplate technic with the same A1 and B test cells. These two studies showed that all the tests from the multiples combinations of the above parameters gave the same results and allowed us to define a range of tolerance for 4 critical physical parameters involved in the antibody screening and blood typing.
ISSN:1246-7820
DOI:10.1016/j.tracli.2006.07.011