Seven-Day Artificial Lung Testing in an In-Parallel Configuration
Background A thoracic artificial lung, the MC3 Biolung, is being developed as a bridge to lung transplantation or as a treatment for acute respiratory insufficiency. Methods The thoracic artificial lung was tested in 10 sheep with the goal of 7 days of respiratory support. The sheep were recovered f...
Gespeichert in:
Veröffentlicht in: | The Annals of thoracic surgery 2007-09, Vol.84 (3), p.988-994 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background A thoracic artificial lung, the MC3 Biolung, is being developed as a bridge to lung transplantation or as a treatment for acute respiratory insufficiency. Methods The thoracic artificial lung was tested in 10 sheep with the goal of 7 days of respiratory support. The sheep were recovered from surgery and monitored awake for 7 days. Hemodynamics, blood gases, blood cell counts, and organ function were recorded, and after 7 days, all sheep were euthanized and necropsied. Results Seven sheep survived the full duration. Cardiac output and mean arterial blood pressure were unchanged, averaging 4.7 ± 0.8 L/min and 98 ± 10 mm Hg, respectively. Arterial oxygen tension and device oxygen transfer rate were also unchanged, averaging 110 ± 26 mm Hg and 97.7 ± 35 mL/min, respectively. Arterial carbon dioxide tension was within normal ranges during the entire experiment, averaging 37.4 ± 3.8 mm Hg. Artificial lung blood flow decreased from 51% ± 14% of cardiac output on day 1 to 30% ± 16% by day 7 because of changes in natural and artificial lung resistance. White blood cell counts were significantly elevated on days 5 and 7, and lastly, kidney and liver function remained normal, although signs of kidney infarction or hemorrhage were noted. Conclusions The thoracic artificial lung is suitable for 7-day attachment, but improvements in blood biocompatibility are warranted. |
---|---|
ISSN: | 0003-4975 1552-6259 |
DOI: | 10.1016/j.athoracsur.2007.03.016 |