A New Device For the Treatment of Thromboembolic Strokes
We present data concerning the design and testing of a new clot-sucking device for possible use in the extraction of blood clots that arise during thromboembolic strokes. Our data indicate that a jellylike mass of 0.02 g (of an artificial clot) can be removed from plastic tubing using this vortex-cr...
Gespeichert in:
Veröffentlicht in: | Journal of stroke and cerebrovascular diseases 2007-07, Vol.16 (4), p.167-172 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We present data concerning the design and testing of a new clot-sucking device for possible use in the extraction of blood clots that arise during thromboembolic strokes. Our data indicate that a jellylike mass of 0.02 g (of an artificial clot) can be removed from plastic tubing using this vortex-creating device, which has an optimal diameter of 1 mm, in conjunction with a 6F arterial catheter that is 110-cm long and using applied suction pressures of about 50 cm Hg. |
---|---|
ISSN: | 1052-3057 1532-8511 |
DOI: | 10.1016/j.jstrokecerebrovasdis.2007.03.003 |