A New Device For the Treatment of Thromboembolic Strokes

We present data concerning the design and testing of a new clot-sucking device for possible use in the extraction of blood clots that arise during thromboembolic strokes. Our data indicate that a jellylike mass of 0.02 g (of an artificial clot) can be removed from plastic tubing using this vortex-cr...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of stroke and cerebrovascular diseases 2007-07, Vol.16 (4), p.167-172
Hauptverfasser: Pearce, Gillian, PhD, BMBCh, BSc, PGCE(HE), Patrick, John H., FRCS, Perkinson, Neil D
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We present data concerning the design and testing of a new clot-sucking device for possible use in the extraction of blood clots that arise during thromboembolic strokes. Our data indicate that a jellylike mass of 0.02 g (of an artificial clot) can be removed from plastic tubing using this vortex-creating device, which has an optimal diameter of 1 mm, in conjunction with a 6F arterial catheter that is 110-cm long and using applied suction pressures of about 50 cm Hg.
ISSN:1052-3057
1532-8511
DOI:10.1016/j.jstrokecerebrovasdis.2007.03.003