Fatigue Fracture of a Proximally Modular, Distally Tapered Fluted Implant with Diaphyseal Fixation

Abstract We report and analyze the causes of a fracture in a proximally modular, distally tapered fluted MP stem in a 48-year-old woman (168 cm, 67 kg) with severe proximal bone deficiency. Evidence of fatigue failure with striations initiated laterally was observed in the laser etching of the tensi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of arthroplasty 2007-08, Vol.22 (5), p.780-783
Hauptverfasser: Buttaro, Martín A., MD, Mayor, Michael B., MD, Van Citters, Douglas, PhD, Piccaluga, Francisco, MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Abstract We report and analyze the causes of a fracture in a proximally modular, distally tapered fluted MP stem in a 48-year-old woman (168 cm, 67 kg) with severe proximal bone deficiency. Evidence of fatigue failure with striations initiated laterally was observed in the laser etching of the tensile aspect of the prosthesis. However, metallurgical analysis suggested that laser engraving did not alter the microstructure of the stem. Stress due to the absence of proximal femoral bone support may have been sufficiently high to put this particular stem at risk for fatigue fracture. This important complication should be addressed when choosing this therapeutic option in cases with substantial proximal femoral bone loss. Strut allograft support should be recommended in such cases.
ISSN:0883-5403
1532-8406