Early experience with remote pressure sensor respiratory plethysmography monitoring sedation in the MR scanner
BACKGROUND AND OBJECTIVE:The importance of monitoring the breathing pattern during sedation of children undergoing magnetic resonance scans is indicated in guidelines, but no appropriate magnetic resonance-compatible devices are available. We report preliminary findings from a technique referred to...
Gespeichert in:
Veröffentlicht in: | European journal of anaesthesiology 2007-09, Vol.24 (9), p.761-769 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | BACKGROUND AND OBJECTIVE:The importance of monitoring the breathing pattern during sedation of children undergoing magnetic resonance scans is indicated in guidelines, but no appropriate magnetic resonance-compatible devices are available. We report preliminary findings from a technique referred to as remote pressure sensor respiratory plethysmography.
METHODS:A data acquisition system was developed, enabling measurement of respiratory rate, plethysmogram amplitude, proportion of inspiratory time over cycle time, thoraco-abdominal phase shift and sigh rate. Correlation between plethysmogram amplitude and tidal volume was investigated on adult volunteers. Twenty-seven children undergoing sedation were monitored with remote pressure sensor respiratory plethysmography, in addition to SPO2 and PetCO2. Differences in monitoring parameters were searched for among three groupspatients who received chloral hydrate only (chloral succeeded, CS group), those who received a supplementation of sodium thiopental (chloral failed, CF group), and those who were sedated with sodium thiopental directly (no chloral, NC group). Correlations were searched for among monitoring parameters, and with total dose of thiopental. The long-term behaviour of respiratory rate, proportion of inspiratory time over cycle time and phase shift was studied.
RESULTS:Plethysmogram amplitude was found to correlate linearly with tidal volume (r > 0.92), with a slope varying up to 22%. While 11% of patients did not tolerate the capnometric probe and readings were discontinuous in 26%, all of them tolerated remote pressure sensor respiratory plethysmography belts. Sighs and non-respiratory movements of the torso could be distinguished on remote pressure sensor respiratory plethysmography waveforms. No significant inter-group differences were found in PetCO2, SPO2, respiratory rate and phase shift. Proportion of inspiratory time over cycle time was higher in the NC group when compared to the CS group (0.497 ± 0.03 vs. 0.463 ± 0.008; P = 0.02), the CF group being characterized by intermediate values (0.480 ± 0.008); when compared to the CS group, sigh rate was lower in the CF group (0.04 ± 0.04 vs. 0.14 ± 0.08; P = 0.04) and in the NC group (0.06 ± 0.05 vs. 0.14 ± 0.08, P = 0.03). A positive correlation was found between total dose of thiopental and proportion of inspiratory time over cycle time, with r = 0.4 and P = 0.04. A large baseline variability in phase shift was found. No long-term trends predi |
---|---|
ISSN: | 0265-0215 1365-2346 |
DOI: | 10.1017/S0265021507000312 |