High performance liquid chromatographic methods for the determination of aripiprazole with ultraviolet detection in rat plasma and brain : Application to the pharmacokinetic study
High performance liquid chromatographic (HPLC) methods were validated for the determination of aripiprazole (OPC-14597, Abilify) in rat plasma and brain. Separation was by Nova-pak phenyl column; flow rate, 1.0 ml/min; mobile phase, acetonitrile-methanol-20 mM sodium sulfate-acetic acid (27:25:48:1,...
Gespeichert in:
Veröffentlicht in: | Journal of chromatography. B, Analytical technologies in the biomedical and life sciences Analytical technologies in the biomedical and life sciences, 2005-07, Vol.821 (1), p.8-14 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | High performance liquid chromatographic (HPLC) methods were validated for the determination of aripiprazole (OPC-14597, Abilify) in rat plasma and brain. Separation was by Nova-pak phenyl column; flow rate, 1.0 ml/min; mobile phase, acetonitrile-methanol-20 mM sodium sulfate-acetic acid (27:25:48:1, v/v/v/v); UV detection at 254 nm. Reproducibility in plasma and brain showed excellent precision (within 7.8 and 10.6%) and accuracy (96.0-102.4% and 99.0-108.7%) with calibration curve ranges 10.0-2000 ng/ml and 30.0-6000 ng/g, respectively. Validated HPLC methods were successfully applied to pharmacokinetic study of aripiprazole in rats, demonstrating brain concentrations after oral administration five times higher than plasma concentrations. |
---|---|
ISSN: | 1570-0232 1873-376X |
DOI: | 10.1016/j.jchromb.2005.03.024 |