Fluted tapered revision femoral stems have been used successfully in Europe. Modularity makes these implants more versatile, but outcomes of modular designs have not been reported. We report a retrospective review of 70 such stems with a mean follow-up of 47 months. Combined metaphyseal/diaphyseal b...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of arthroplasty 2006-04, Vol.21 (3), p.372-380
Hauptverfasser: McInnis, Douglas P, Horne, Geoffrey, Devane, Peter A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Fluted tapered revision femoral stems have been used successfully in Europe. Modularity makes these implants more versatile, but outcomes of modular designs have not been reported. We report a retrospective review of 70 such stems with a mean follow-up of 47 months. Combined metaphyseal/diaphyseal bone loss was present preoperatively in 36 (51%) of 70 hips. Three hips (4.3%) were re-revised or in need of re-revision, and worst-case survival was 87%. The mean postoperative patient-assessed Oxford Hip Score was 21.1. Restoration of proximal bone was noted in 56%. Complications included mean subsidence of 9.9 mm, dislocation in 7 (10%) of 70 hips, and fracture or cortical perforation in 17 (24.2%) of 70. These results are equivalent to, or better than, reported results of other types of cementless revision femoral stems.
ISSN:0883-5403
DOI:10.1016/j.arth.2005.08.022