Technique for safe removal of an aortic endograft with suprarenal fixation
Complete removal of an aortic endograft with suprarenal fixation is difficult. We report the use of a simple device to extract a Zenith endograft. This device is made by cutting off the tip of the cylinder of a 20-mL syringe and rounding off the edges. The main body of the graft is resheathed by adv...
Gespeichert in:
Veröffentlicht in: | Journal of vascular surgery 2006-04, Vol.43 (4), p.855-857 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Complete removal of an aortic endograft with suprarenal fixation is difficult. We report the use of a simple device to extract a Zenith endograft. This device is made by cutting off the tip of the cylinder of a 20-mL syringe and rounding off the edges. The main body of the graft is resheathed by advancing the cylinder cranially while keeping the graft in a stable position. This way, the graft collapses and the hooks are withdrawn without tearing the wall of the aorta. The supraceliac clamping time in our patient was less than 2 minutes. This technique offers a safe and fast solution to the potentially hazardous removal of an aortic endograft with suprarenal fixation. |
---|---|