Finite element analysis of covered microstents

Currently available neuroendovascular devices are inadequate for effective treatment of many wide-necked or fusiform intracranial aneurysms and intracranial carotid-cavernous fistulae (CCF). Placing a covered microstent across the intracranial aneurysm neck and CCF rent could restore normal vessel m...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of biomechanics 2005-06, Vol.38 (6), p.1221-1227
Hauptverfasser: Gu, Linxia, Santra, Swadeshmukul, Mericle, Robert A., Kumar, Ashok V.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Currently available neuroendovascular devices are inadequate for effective treatment of many wide-necked or fusiform intracranial aneurysms and intracranial carotid-cavernous fistulae (CCF). Placing a covered microstent across the intracranial aneurysm neck and CCF rent could restore normal vessel morphology by preventing blood flow into the aneurysm lumen or CCF rent. To fabricate covered microstents, our research group has developed highly flexible ultra thin (∼150μm) silicone coverings and elastomerically captured them onto commercially available metal stents without stitching. Preliminary in vivo studies were conducted by placing these covered microstents in the common carotid artery of rabbits. The feasibility of using covered stents was demonstrated. However, the cover affected the deployment pressure and the stents failed occasionally during deployment due to tearing of the cover. Appropriate modeling of covered stents will assist in designing suitable coverings, and help to reduce the failure rate of covered microstents. The purpose of this study is to use the finite element method to determine the mechanical properties of the covered microstent and investigate the effects of the covering on the mechanical behavior of the covered microstent. Variations in the mechanical properties of the covered microstent such as deployment pressure, elastic recoil and longitudinal shortening due to change in thickness and material properties of the cover have been investigated. This work is also important for custom design of covered microstents such as adding cutout holes to save adjacent perforating arteries.
ISSN:0021-9290
1873-2380
DOI:10.1016/j.jbiomech.2004.06.008