Contribution of early systole to total antegrade flow volume in breath-hold phase-contrast flow measurements
To evaluate the contribution of early systole for the assessment of antegrade aortic flow volume by breath-hold velocity-encoded magnetic resonance (MR) flow measurements. Expiratory breath-hold fast low-angle shot (FLASH) phase-contrast flow measurements (temporal resolution tRes 61 msec, shared ph...
Gespeichert in:
Veröffentlicht in: | RöFo : Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebende Verfahren 2005-05, Vol.177 (5), p.637-645 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | ger |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To evaluate the contribution of early systole for the assessment of antegrade aortic flow volume by breath-hold velocity-encoded magnetic resonance (MR) flow measurements.
Expiratory breath-hold fast low-angle shot (FLASH) phase-contrast flow measurements (temporal resolution tRes 61 msec, shared phases) perpendicular to the proximal ascending aorta and short axis true fast imaging with steady-state precession (TrueFISP) cine MR ventriculometry (tRes 34.5 msec) were performed in ten healthy male volunteers on a 1.5 T MR system (Sonata, Siemens Medical Solutions). Antegrade aortic flow volume (AFV) and left ventricular stroke volume (LV-SV) were evaluated using Argus Ventricular Function and Argus Flow Software, version MR 2002B (Siemens Medical Solutions). A beta release of Argus Flow MR 2004A allowed interpolation of the flow up-slope during early systole to the preceding R-wave trigger. The respective intraindividual median differences between the AFV of each flow evaluation and LV-SV as well as between both AFV measurements were calculated and compared using the sign test for paired samples.
Non-interpolated AFV significantly deviated from LV-SV (p = 0.006), underestimating the latter by 13.1 mL (13 %). Interpolating aortic flow during early systole significantly increased AFV by 10.8 mL (13 %) compared to the flow evaluation which did not take early systole into account (p = 0.006). AFV with interpolation of early systolic flow agreed well with LV-SV (median difference - 3.0 mL or - 3 %, respectively), and no significant difference between these measurements was found (p = 1.0).
Flow during early systole contributes substantially to total forward flow volume in the ascending aorta. Interpolation of the early systolic up-slope is therefore recommended for the evaluation of breath-hold phase-contrast flow measurements. |
---|---|
ISSN: | 1438-9029 |