Performance of the DNA-Citoliq liquid-based cytology system compared with conventional smears
Objective: To evaluate the performance of a new, manual, simplified liquid‐based system, DNA‐Citoliq (Digene Brasil), employed under routine conditions as compared to conventional smears collected from six collaborating private laboratories. Methods: A panel of cytopathologists, who served as the go...
Gespeichert in:
Veröffentlicht in: | Cytopathology (Oxford) 2006-04, Vol.17 (2), p.86-93 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objective: To evaluate the performance of a new, manual, simplified liquid‐based system, DNA‐Citoliq (Digene Brasil), employed under routine conditions as compared to conventional smears collected from six collaborating private laboratories.
Methods: A panel of cytopathologists, who served as the gold standard diagnosis, adjudicated discordant opinions.
Results: Of 3206 pairs of slides considered valid for comparison, there were 3008 in full agreement (93.8%), 112 (3.5%) with one diagnostic category discrepancies, and 86 (2.7%) discordant cases. Among the 288 borderline+ by either method, DNA‐Citoliq detected abnormalities in 243 (84.4%), and conventional smears (CS) detected abnormalities in 178 (61.8%) (McNemar test, P |
---|---|
ISSN: | 0956-5507 1365-2303 |