In vitro validation of phase-contrast flow measurements at 3 T in comparison to 1.5 T: Precision, accuracy, and signal-to-noise ratios

Purpose To evaluate the signal‐to‐noise ratio (SNR), precision, and accuracy of phase‐contrast flow measurements at 3 T with the help of an in vitro model and to compare the results with data from two 1.5‐T scanners. Materials and Methods Using an identical setup of a laminar flow model and sequence...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of magnetic resonance imaging 2005-05, Vol.21 (5), p.604-610
Hauptverfasser: Lotz, Joachim, Döker, Rolf, Noeske, Ralph, Schüttert, Meike, Felix, Roland, Galanski, Michael, Gutberlet, Matthias, Meyer, Gerd Peter
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose To evaluate the signal‐to‐noise ratio (SNR), precision, and accuracy of phase‐contrast flow measurements at 3 T with the help of an in vitro model and to compare the results with data from two 1.5‐T scanners. Materials and Methods Using an identical setup of a laminar flow model and sequence parameters, measurements were done at one 3‐T and at two 1.5‐T systems. Precision, accuracy, and SNR were obtained for velocity encodings ranging from 55 up to 550 cm−1. SNRs were calculated from the magnitude as well as the flow encoded images. Results Precision and accuracy for the in vitro flow model were similarly high in all scanners with no significant difference. For velocity encodings from 55 cm−1 up to 550 cm−1, the SNR in magnitude as well as phase encoded images of the 3‐T measurements was approximately 2.5 times higher than the SNR obtained from the two 1.5‐T systems. Conclusion Even without optimization for the 3‐T environment, flow measurements show the same high accuracy and precision as is known from clinical 1.5‐T scanners. The superior SNR at 3 T will allow further improvements in temporal and spatial resolution. This will be of interest for small‐size vessels like coronary arteries or for slow diastolic flow patterns. J. Magn. Reson. Imaging 2005;21:604–610. © 2005 Wiley‐Liss, Inc.
ISSN:1053-1807
1522-2586
DOI:10.1002/jmri.20275