A novel device for measuring long-term oxygen therapy adherence : A preliminary validation
Current methods for measuring patient adherence to long-term oxygen therapy fail to measure the actual amount of time the patient is inhaling oxygen and the pattern of oxygen use within the day. We have developed a novel oxygen-adherence monitor to address these limitations, and this report introduc...
Gespeichert in:
Veröffentlicht in: | Respiratory care 2006-03, Vol.51 (3), p.266-271 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Current methods for measuring patient adherence to long-term oxygen therapy fail to measure the actual amount of time the patient is inhaling oxygen and the pattern of oxygen use within the day. We have developed a novel oxygen-adherence monitor to address these limitations, and this report introduces the monitor and provides preliminary data validating its use.
This battery-powered monitor attaches to the oxygen source and detects respiratory-related pressure fluctuations transmitted through the nasal cannula. The monitor takes a measurement over a 25-second period, at 4-min intervals. It detects and stores data on 4 different states that describe the patient's actual use of the oxygen source and nasal cannula: source-off/cannula-off, source-off/cannula-on, source-on/cannula-off, and source-on/cannula-on. We studied the monitor's performance with 10 patients with chronic obstructive pulmonary disease, during a directly-observed sequence of using and not using supplemental oxygen via nasal cannula, while sitting and walking.
The monitor correctly detected 122 out of 129 measurements among all participants, yielding a 95\% detection accuracy.
A monitor that objectively measures oxygen inhalation, rather than oxygen expenditure, may help improve the management of patients on long-term oxygen therapy. |
---|---|