Sterility testing of platelet concentrates prepared from deliberately infected blood donations

BACKGROUND:  In general the bacterial count in freshly donated blood is low and even lower in the corresponding platelet concentrates (PCs). By use of flow cytometry (FACS) for sterility testing, the reliability of early versus later sampling times was evaluated. STUDY DESIGN AND METHODS:  Blood don...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Transfusion (Philadelphia, Pa.) Pa.), 2006-03, Vol.46 (3), p.486-491
Hauptverfasser: Mohr, Harald, Lambrecht, Bernd, Bayer, Anette, Spengler, Hans-Peter, Nicol, Sven-Boris, Montag, Thomas, Müller, Thomas H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:BACKGROUND:  In general the bacterial count in freshly donated blood is low and even lower in the corresponding platelet concentrates (PCs). By use of flow cytometry (FACS) for sterility testing, the reliability of early versus later sampling times was evaluated. STUDY DESIGN AND METHODS:  Blood donations were spiked with various numbers of Staphylococcus epidermidis, Staphylococcus aureus, Bacillus cereus, and Klebsiella pneumoniae. The corresponding PCs were prepared by the buffy‐coat method and stored at 22°C. A 20‐mL sample was collected from each PC directly after preparation and after 8 hours. Samples were stored at 35°C. Sterility testing of both PCs and samples was by FACS analysis at different time points. RESULTS:  All stored PCs were found positive by FACS analysis, with detection times ranging between 8 and 24 hours (K. pneumoniae, B. cereus), 8 and 91 hours (S. aureus), and 144 hours (S. epidermidis). In the samples incubated at 35°C, bacteria were detected after 8 to 19 hours (K. pneumoniae, B. cereus), 8 to 67 hours (S. aureus), and 19 to 43 hours (S. epidermidis). Some of the samples did not contain bacteria. CONCLUSION:  Detection times for slow‐growing bacteria are significantly shortened when PC samples are incubated at 35°C: The numbers of bacteria in freshly prepared PCs may, however, be so low that the samples drawn for sterility testing do not contain a single bacterium. Our results do not support a shortening of the 24‐hour or greater sampling time recommended by the manufacturers of established test systems, because also for consistent detection by FACS, bacteria need to grow in the PCs to sufficient numbers.
ISSN:0041-1132
1537-2995
DOI:10.1111/j.1537-2995.2006.00747.x