64-slice multidetector coronary CT angiography: in vitro evaluation of 68 different stents
The purpose of this study was to test a large sample of different coronary artery stents using four image reconstruction approaches with respect to lumen visualization, lumen attenuation, and image noise in 64-slice multidetector-row computed tomography (MDCT) in vitro and to provide a catalogue of...
Gespeichert in:
Veröffentlicht in: | European radiology 2006-04, Vol.16 (4), p.818-826 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The purpose of this study was to test a large sample of different coronary artery stents using four image reconstruction approaches with respect to lumen visualization, lumen attenuation, and image noise in 64-slice multidetector-row computed tomography (MDCT) in vitro and to provide a catalogue of currently used coronary artery stents when imaged with state-of the-art MDCT. We examined 68 different coronary artery stents (57 stainless steel, four cobalt-chromium, one cobalt-alloy, two nitinol, four tantalum) in a coronary artery phantom (vessel diameter 3 mm, intravascular attenuation 250 HU, extravascular density -70). Stents were imaged in axial orientation with standard parameters: 32 x 0.6 collimation, pitch 0.24, 680 mAs, 120 kV, rotation time 0.37 s. Four different image reconstructions were obtained with varying convolution kernels and section thicknesses: (1) soft, 0.6 mm, (2) soft, 0.75, (3) medium soft, 0.6, and (4) stent-optimized sharp, 0.6. To evaluate visualization characteristics of of the stent, the lumen diameter, intraluminal density and noise were measured. The high-resolution kernel offered significantly better average lumen visualization (57% +/-10%) and more realistic lumen attenuation (222 HU +/-66 HU) at the expense of increased noise (15.3 HU +/-3.7 HU) compared with the soft and medium-soft CT angiography (CTA) protocol (p |
---|---|
ISSN: | 0938-7994 1432-1084 |
DOI: | 10.1007/s00330-005-0062-8 |