Between January 1987 and December 1990, 221 Zweymüller cementless total hip arthroplasties were performed in 211 patients with idiopathic osteoarthritis. A total of 136 patients (142 prostheses) were evaluated at a mean follow-up of 134 months (SD 9.5). The study group consisted of 78 Hochgezogen an...
Gespeichert in:
Veröffentlicht in: | The Journal of arthroplasty 2005-04, Vol.20 (3), p.362-368 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Between January 1987 and December 1990, 221 Zweymüller cementless total hip arthroplasties were performed in 211 patients with idiopathic osteoarthritis. A total of 136 patients (142 prostheses) were evaluated at a mean follow-up of 134 months (SD 9.5). The study group consisted of 78 Hochgezogen and 64 Stepless stem prostheses, all with a threaded titanium cup and ceramic head. No clinical and radiological differences were found between the 2 stem prostheses. Seven cups had been revised because of aseptic loosening; 17 cups showed radiolucent lines, osteolysis, or migration. Mean linear polyethylene wear of 105 (74%) cups was 0.46 mm (SD 0.27), with an annual wear of 0.04 mm (SD 0.02). Wear did not correlate with pain, cup migration, radiolucent lines, or osteolysis. Cumulative survival was 96%. Zweymüller cementless total hip arthroplasty showed good midterm results. |
---|---|