The Long-Term Wear of Retrieved McKee-Farrar Metal-on-Metal Total Hip Prostheses

Twenty-four cobalt-chrome alloy McKee-Farrar matching acetabular and femoral components were retrieved at revision total hip arthroplasty. The average time in situ was 16 years. Wear and loss of sphericity was very low. Polishing wear (type 1), fine abrasive (type 2), multidirectional dull abrasive...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of arthroplasty 2005-04, Vol.20 (3), p.350-357
Hauptverfasser: Howie, Donald W., McCalden, Richard W., Nawana, Namal S., Costi, Kerry, Pearcy, Mark J., Subramanian, Chinnia
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Twenty-four cobalt-chrome alloy McKee-Farrar matching acetabular and femoral components were retrieved at revision total hip arthroplasty. The average time in situ was 16 years. Wear and loss of sphericity was very low. Polishing wear (type 1), fine abrasive (type 2), multidirectional dull abrasive (type 3), and unidirectional dull abrasive wear (type 4) of the articulating surfaces were identified. The mean percent area of femoral heads occupied by types 2, 3, and 4 wear was 18%, 5%, and 2%, respectively. There was no association between the type and distribution of wear and the time in situ. Impingement damage was evident on 9 implant pairs. The extent and types of wear described in this paper will be useful when analyzing the patterns of surface damage of newer designs of metal-on-metal articulations.
ISSN:0883-5403
1532-8406
DOI:10.1016/j.arth.2004.09.028