An Ill Wind‐Iatrogenic Air Embolus Around Pacing Leads During Defibrillator Implant with Coexisting Pulmonary Fibrosis

We present a case of air embolism during biventricular defibrillator implantation in a 58‐year‐old man with pulmonary fibrosis. To prevent air entry after the dilator was removed, the subclavian venous sheath was covered before the lead was introduced. Air embolism occurred during inspiration around...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 2006-02, Vol.29 (2), p.213-214
Hauptverfasser: NINIO, DANIEL M, HII, JOHN T
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We present a case of air embolism during biventricular defibrillator implantation in a 58‐year‐old man with pulmonary fibrosis. To prevent air entry after the dilator was removed, the subclavian venous sheath was covered before the lead was introduced. Air embolism occurred during inspiration around the lead body in the appropriate sized sheath but without a hemostatic valve. Extraordinary swings in intrathoracic pressure due to noncompliant lungs may have contributed to this unusual complication.