Performance of MicroScan WalkAway and Vitek 2 for Detection of Oxacillin Resistance in a Set of Methicillin-Resistant Staphylococcus aureus Isolates with Diverse Genetic Backgrounds
Of 104 genotypically diverse methicillin-resistant Staphylococcus aureus (MRSA) isolates tested with the MicroScan WalkAway (Pos MIC 24 panel) and Vitek 2 (AST-P549 card) systems, 7 and 6 isolates, respectively, showed an oxacillin MIC of [less-than or equal to]2mg/liter. Most of these MRSA isolates...
Gespeichert in:
Veröffentlicht in: | Journal of Clinical Microbiology 2009-08, Vol.47 (8), p.2623-2625 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Of 104 genotypically diverse methicillin-resistant Staphylococcus aureus (MRSA) isolates tested with the MicroScan WalkAway (Pos MIC 24 panel) and Vitek 2 (AST-P549 card) systems, 7 and 6 isolates, respectively, showed an oxacillin MIC of [less-than or equal to]2mg/liter. Most of these MRSA isolates were community acquired. However, if the cefoxitin screen of AST-P549 was also considered, MRSA detection failed for only one isolate. |
---|---|
ISSN: | 0095-1137 1098-660X |
DOI: | 10.1128/JCM.02112-08 |