Modular Ulnar Head Decoupling Force: Case Report
Cobalt-chrome modular distal ulnar head replacement arthroplasty is a surgical option to restore stability to the distal radioulnar joint rendered unstable by hemi-resection arthroplasty or a total resection arthroplasty. However, the revision of dislocated modular cobalt-chrome ulnar head implants...
Gespeichert in:
Veröffentlicht in: | The Journal of hand surgery (American ed.) 2009-07, Vol.34 (6), p.1063-1065 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Cobalt-chrome modular distal ulnar head replacement arthroplasty is a surgical option to restore stability to the distal radioulnar joint rendered unstable by hemi-resection arthroplasty or a total resection arthroplasty. However, the revision of dislocated modular cobalt-chrome ulnar head implants may pose an important intraoperative challenge. The Morse-taper disassembly force of modular ulnar head implants is not available in the current published literature. We present a case in which tremendous difficulty was encountered while revising a dislocated modular cobalt-chrome distal ulnar head implant. The mean Morse-taper disassembly force of the retrieved modular cobalt-chrome implant was 2958 N ± 1272. At nearly 4.5 times the average body weight, the modular ulnar head Morse-taper disassembly strength presented a formidable force to overcome intraoperatively. |
---|---|
ISSN: | 0363-5023 1531-6564 |
DOI: | 10.1016/j.jhsa.2009.02.013 |