Narrowband Magnetic Particle Imaging
The magnetic particle imaging (MPI) method directly images the magnetization of super-paramagnetic iron oxide (SPIO) nanoparticles, which are contrast agents commonly used in magnetic resonance imaging (MRI). MPI, as originally envisioned, requires a high-bandwidth receiver coil and preamplifier, wh...
Gespeichert in:
Veröffentlicht in: | IEEE transactions on medical imaging 2009-08, Vol.28 (8), p.1231-1237 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The magnetic particle imaging (MPI) method directly images the magnetization of super-paramagnetic iron oxide (SPIO) nanoparticles, which are contrast agents commonly used in magnetic resonance imaging (MRI). MPI, as originally envisioned, requires a high-bandwidth receiver coil and preamplifier, which are difficult to optimally noise match. This paper introduces Narrowband MPI, which dramatically reduces bandwidth requirements and increases the signal-to-noise ratio for a fixed specific absorption rate. We employ a two-tone excitation (called intermodulation) that can be tailored for a high- Q , narrowband receiver coil. We then demonstrate a new MPI instrument capable of full 3-D tomographic imaging of SPIO particles by imaging acrylic and tissue phantoms. |
---|---|
ISSN: | 0278-0062 1558-254X |