Support adapter for radiofrequency ablation probe: experimental study
Background: The purpose of the present study was to experimentally evaluate a new support adapter, which was created in order to insure a more complete contact between the ablation probe and a beating heart. Methods: Support sleeve adapters for the Cobra Cooled probe were fashioned using a semi‐ri...
Gespeichert in:
Veröffentlicht in: | ANZ journal of surgery 2005-03, Vol.75 (3), p.157-159 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background: The purpose of the present study was to experimentally evaluate a new support adapter, which was created in order to insure a more complete contact between the ablation probe and a beating heart.
Methods: Support sleeve adapters for the Cobra Cooled probe were fashioned using a semi‐rigid plastic. Forty epicardial radiofrequency ablation (RFA) lesions were created on both atria under the condition of a beating heart in six pigs weighing 26−30 kg. A Cobra Cooled surgical probe was used with continuous internal irrigation of a saline solution (500 mL/h). The ablation temperature was fixed at 80°C and the duration of the RFA in each case was 20, 30 and 60 s.
Results: A support adapter ensured a more complete epicardial contact between the probe and a beating heart. An RFA probe supported by the adapter delivered greater power to the tissue. The histopathological transmural coagulation was obtained in 82% and 72% in RFA lesions with and without the adaptor, respectively.
Conclusion: A new adapter was effective in the epicardial RFA under the condition of a beating heart. |
---|---|
ISSN: | 1445-1433 1445-2197 |