Cementation of a polyethylene liner into a metal acetabular shell: a biomechanical study

Cementation of a liner into a well-fixed acetabular shell is common in revision hip arthroplasty. We compare the biomechanical strengths of cemented liners with standard locked liners. Fifty polyethylene liners were inserted into acetabular shells using the standard locking mechanism or 1 of 2 cemen...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The Journal of arthroplasty 2009-08, Vol.24 (5), p.775-782
Hauptverfasser: Hofmann, Aaron A, Prince, Edward J, Drake, F Thurston, Hunt, Kenneth J
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Cementation of a liner into a well-fixed acetabular shell is common in revision hip arthroplasty. We compare the biomechanical strengths of cemented liners with standard locked liners. Fifty polyethylene liners were inserted into acetabular shells using the standard locking mechanism or 1 of 2 cement types then loaded to failure by torsion or lever-out testing. Lever-out testing showed that all cemented liners failed at similar loads to standard locked liners. With torsion testing, cemented liners failed at significantly higher loads than standard locked liners; roughening the liner increased load to failure. Cementation of an acetabular liner into a metal shell is safe and strong and a good alternative to metal shell replacement. Saw roughening of the polyethylene liner strengthens the poly-cement interface.
ISSN:0883-5403
1532-8406