Varying Effects of Corneal Thickness on Intraocular Pressure Measurements with Different Tonometers
Aims To compare the intraocular pressure (IOP) measurements using the Goldmann applanation tonometer (GAT), noncontact tonometer (NCT), and ocular blood flow tonometer (OBFT), and to evaluate the effects of varying central corneal thickness (CCT) on the readings. Methods Ultrasound pachymetry and to...
Gespeichert in:
Veröffentlicht in: | Eye (London) 2005-03, Vol.19 (3), p.327-332 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Aims
To compare the intraocular pressure (IOP) measurements using the Goldmann applanation tonometer (GAT), noncontact tonometer (NCT), and ocular blood flow tonometer (OBFT), and to evaluate the effects of varying central corneal thickness (CCT) on the readings.
Methods
Ultrasound pachymetry and tonometry were performed on 170 eyes. Using the corrected GAT values that took CCT into account as the standard, we calculated the NCT and OBFT measurement errors related to the CCT variable by a linear regression model. Group comparisons were performed with the
χ
2
test and one-way ANOVA test. The correlation between the various tonometer measurements was analysed by Pearson's correlation method.
Results
Both the NCT (
r
=0.872,
P |
---|---|
ISSN: | 0950-222X 1476-5454 |
DOI: | 10.1038/sj.eye.6701458 |