Varying Effects of Corneal Thickness on Intraocular Pressure Measurements with Different Tonometers

Aims To compare the intraocular pressure (IOP) measurements using the Goldmann applanation tonometer (GAT), noncontact tonometer (NCT), and ocular blood flow tonometer (OBFT), and to evaluate the effects of varying central corneal thickness (CCT) on the readings. Methods Ultrasound pachymetry and to...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Eye (London) 2005-03, Vol.19 (3), p.327-332
Hauptverfasser: Ko, Y-C, Liu, CJ-l, Hsu, W-M
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Aims To compare the intraocular pressure (IOP) measurements using the Goldmann applanation tonometer (GAT), noncontact tonometer (NCT), and ocular blood flow tonometer (OBFT), and to evaluate the effects of varying central corneal thickness (CCT) on the readings. Methods Ultrasound pachymetry and tonometry were performed on 170 eyes. Using the corrected GAT values that took CCT into account as the standard, we calculated the NCT and OBFT measurement errors related to the CCT variable by a linear regression model. Group comparisons were performed with the χ 2 test and one-way ANOVA test. The correlation between the various tonometer measurements was analysed by Pearson's correlation method. Results Both the NCT ( r =0.872, P
ISSN:0950-222X
1476-5454
DOI:10.1038/sj.eye.6701458