Calibration of the photon component of 198Au stents

198Au has promising characteristics for radioactive stent material, having properties as a mixed β-particle and γ emitter. Calibration of these radioactive stents is required to provide accurate clinical dosimetry. We have developed an electroplating technique to incorporate stable gold onto stents...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Brachytherapy 2005, Vol.4 (1), p.51-58
Hauptverfasser: Culberson, Wesley S., DeWerd, Larry A., Thomadsen, Bruce R., Micka, John A., Bohm, Tim D.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:198Au has promising characteristics for radioactive stent material, having properties as a mixed β-particle and γ emitter. Calibration of these radioactive stents is required to provide accurate clinical dosimetry. We have developed an electroplating technique to incorporate stable gold onto stents followed by activation to 198Au in the University of Wisconsin nuclear reactor. The calibration method is a modification of the NIST traceable, in-air calibration technique for high-dose-rate (HDR) 192Ir sources. The air-kerma strength of HDR and low-dose-rate (LDR) sources was measured for proof of principle and found to agree to within 3% of values obtained with other NIST traceable calibration techniques. The photon component of two 198Au radioactive stents was measured over a period of 3 days. The air-kerma strength of HDR and LDR sources was measured for proof of principle and found to agree to within 3% of values obtained with other NIST traceable calibration techniques.
ISSN:1538-4721
1873-1449
DOI:10.1016/j.brachy.2004.11.004