Reliability of cardiorespiratory measurements with a new ergospirometer during intense treadmill exercise in Thoroughbred horses

This study investigated the reliability of measurements with a new equine ergospirometer (Quadflow). Heart rate and blood lactate responses during exercise in horses wearing the Quadflow and an open flow mask were also compared. The mean percentage error of the oxygen uptake measurements was 8.2% (r...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The veterinary journal (1997) 2005-03, Vol.169 (2), p.223-231
Hauptverfasser: Curtis, R.A., Kusano, K., Evans, D.L., Lovell, N.H., Hodgson, D.R.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:This study investigated the reliability of measurements with a new equine ergospirometer (Quadflow). Heart rate and blood lactate responses during exercise in horses wearing the Quadflow and an open flow mask were also compared. The mean percentage error of the oxygen uptake measurements was 8.2% (range 2.1–12.5%). Percent error for peak expiratory flow rates ranged from 6.1% to 9.4 %, and for minute ventilation from 2.5% to 7.4%. The coefficients of variation of the means of four measurements in two horses exercising continuously at 9.0 m/s were