Measurement of Macular Pigment: Raman Spectroscopy versus Heterochromatic Flicker Photometry
There are several techniques for measuring macular pigment (MP) in vivo, of which Raman spectroscopy (RS) is a recently developed objective This study reports the reproducibility, test-retest variability, and validity of RS MP readings, by comparing them with heterochromatic flicker photometry (HFP)...
Gespeichert in:
Veröffentlicht in: | Investigative ophthalmology & visual science 2005-03, Vol.46 (3), p.1023-1032 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | There are several techniques for measuring macular pigment (MP) in vivo, of which Raman spectroscopy (RS) is a recently developed objective
This study reports the reproducibility, test-retest variability, and validity of RS MP readings, by comparing them with heterochromatic flicker photometry (HFP).
MP was measured with HFP and RS in 120 healthy subjects, and the latter technique was also used on two separate occasions in a sample of 20 subjects to investigate the intersessional variability of readings. Intrasessional reproducibility of RS MP measurements was also calculated. In addition, serum concentrations of lutein (L) and zeaxanthin (Z) were measured and correlated with both RS and HFP MP readings.
Mean (+/-SD) MP in the right eye was 0.279 +/- 0.145 and 0.319 +/- 0.155 with RS and HFP, respectively. The differences between corresponding MP readings taken on RS and HFP lay within the Bland-Altman 95% limits of agreement for the two instruments in 93.6% and 94.4% of cases in the right and left eyes, respectively. Intrasessional reproducibility of RS readings, expressed as the coefficient of variation, was 8.42% +/- 7.12%. Ninety-five percent of MP readings taken with RS on two separate occasions lay within the 95% limits of agreement for the two sessions. A positive, but insignificant, relationship was observed between RS and HFP MP readings and serum concentrations of L and Z (RS, P = 0.356; HFP, P = 0.540).
RS, an objective method of measuring MP levels in vivo, exhibits acceptable reproducibility and test-retest variability. The results demonstrated good correlation between RS and HFP measurements of MP, thus authenticating RS against a validated psychophysical technique of measuring MP. However, investigators should use only one of these instruments for the duration of any given study because of differences in the scientific rationale, and the factors that influence RS and HFP measurements of MP. |
---|---|