Six years' experience of using the BacT/ALERT system to screen all platelet concentrates, and additional testing of outdated platelet concentrates to estimate the frequency of false-negative results
Background and Objectives Approximately 1 in every 2000 units of platelets is contaminated with bacteria. The BacT/ALERT automated blood culture system can be used to screen platelet concentrates (PCs) for bacterial contamination. Materials and Methods Data were collected from May 1998 until May 2...
Gespeichert in:
Veröffentlicht in: | Vox sanguinis 2005-02, Vol.88 (2), p.93-97 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background and Objectives Approximately 1 in every 2000 units of platelets is contaminated with bacteria. The BacT/ALERT automated blood culture system can be used to screen platelet concentrates (PCs) for bacterial contamination.
Materials and Methods Data were collected from May 1998 until May 2004. The number of PCs tested during this period was 36 896, most of which were produced from pools of four buffy‐coats. On the day following blood collection or platelet apheresis, a 5–10 ml sample of the PC was aseptically transferred to a BacT/ALERT culture bottle for detection of aerobic bacteria. The sample was monitored for bacterial growth during the entire storage period of the PC (6·5 days). When a positive signal was generated, the culture bottle, the PC and the erythrocyte concentrates were tested for bacterial growth. In order to determine the frequency of false‐negative BacT/ALERT signals, 1061 outdated PCs were tested during the period from May 2002 to May 2004.
Results Eighty‐eight positive signals were detected by the BacT/ALERT system, of which 12 were interpreted as truly positive. Fourteen signals were interpreted as truly false positive. Thirty‐three signals were interpreted to be probably false positive. Two of 1061 outdated units tested positive, and Bacillus spp. and Staphylococcus epidermidis, respectively, were isolated from these PCs.
Conclusions Between 0·03% and 0·12% of the PCs were contaminated with bacteria. BacT/ALERT is an efficient tool for monitoring PCs for bacterial contamination; however, it is important to realize that false‐negative results may occur. |
---|---|