Computed tomography during cardiopulmonary resuscitation using automated chest compression devices—an initial study
The purpose of the study was to evaluate both CT image quality in a phantom study and feasibility in an initial case series using automated chest compression (A-CC) devices for cardiopulmonary resuscitation (CPR). Multidetector CT (MDCT) of a chest/heart phantom (Thorax-CCI, QRM, Germany) was perfor...
Gespeichert in:
Veröffentlicht in: | European radiology 2009-08, Vol.19 (8), p.1857-1866 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The purpose of the study was to evaluate both CT image quality in a phantom study and feasibility in an initial case series using automated chest compression (A-CC) devices for cardiopulmonary resuscitation (CPR). Multidetector CT (MDCT) of a chest/heart phantom (Thorax-CCI, QRM, Germany) was performed with identical protocols of the phantom alone (S), the phantom together with two different A-CC devices (A: AutoPulse, Zoll, Germany; L: LUCAS, Jolife, Sweden), and the phantom with a LUCAS baseplate, but without the compression unit (L-bp). Nine radiologists evaluated image noise quantitatively (n = 244 regions, Student’s t-test) and also rated image quality subjectively (1–excellent to 6–inadequate, Mann-Whitney U-test). Additionally, three patients during prolonged CPR underwent CT with A-CC devices. Mean image noise of S was increased by 1.21 using L-bp, by 3.62 using A, and by 5.94 using L (p |
---|---|